IJCRR - 14(4), February, 2022
Pages: 63-68
Date of Publication: 15-Feb-2022
Print Article
Download XML Download PDF
A Specific LC-ESI-MS/MS Method Development and Validation for the Quantification of Saquinavir in Biological Matrices
Author: Bandaru Anil Kumar, Bomma Ramesh
Category: Healthcare
Abstract:Introduction: Saquinavir is the first protease inhibitor. Proteases are enzymes that cleave protein molecules into smaller fragments. No analytical procedure was reported for the estimation of saquinavir by LC-MS/MS. Aims: The objective of the present research was to develop a specific liquid chromatography-electrospray ionization tandem mass spectrometric technique for the quantitation of saquinavir in biological matrices. Methodology: Chromatographic elution was attained thru a Thermo Hypersil ODS stationary column having the dimensions of 50 × 4.6 mm and particle size of 2.4 \?m. Isocratic elution was processed with methanol and 0.1%V/V formic acid in the ratio of 90:10 V/V as mobile phase with flow rate of 0.50 ml/min. Liquid-liquid extraction was performed for drug and internal standard isolation with an ethyl acetate solvent. Parent and productions were monitored at m/z 671.3 → 654.3 for saquinavir and 614.3 → 596.3 for indinavir internal standardon multiple reaction monitoring. Result: Linearity graph of drug was rectilinear in concentration over 260.4. to 10416ng/ml having r2(correlation coefficient) value more than 0.999. Percentage of RSD findings were ≤5.3% for inter and intraday accuracy and precision. This procedure has good recoveries and %recovery findings of lower quality control (LQC), median quality control (MQC) and higher quality control (HQC) samples were 94.6%, 92.4 and 104.2% respectively. Conclusion: Saquinavir has more stability for longer time when subjected for different stability environments and the technique was effectively relevant to routine analysis of saquinavir in biological matrix.
Keywords: Protease inhibitor, Saquinavir, LC-ESI-MS/MS, FDA guidelines, Specificity, Linearity
Full Text:
INTRODUCTION
Human immunodeficiency virus (HIV) infection and its clinical syndrome acquired immunodeficiency syndrome (AIDS) continue to be a major health problem worldwide. The development of highly effective antiretroviral chemotherapy for AIDS is found very useful in curbing the pandemic of this disease.1,2 The drug is sold as Fortovase and Invirase brand names and it acts as an antiretroviral agent utilized along with other medicines to prevent or treat HIV/AIDS. Normally it is combined with ritonavir/lopinavir or ritonavir to improve its outcome. Saquinavir is the first protease inhibitor. Proteases are enzymes that cleave protein molecules into smaller fragments.3
The bioavailability of saquinavir increases with ritonavir (PI) in dosage forms which was observed clinically. It will benefit to the patients for taking less dose of saquinavir and also for maintaining adequate plasma levels of saquinavir to produce effective against HIV. In this connection, it was observed that ritonavir prevents isozyme cytochromeP450 3A4, which metabolizes the saquinavir into its inactive metabolite.4
Drug bounds to the effective part of viral protease and inhibits the breakdown of viral polyprotein components, inhibiting the maturation of virus. Saquinavir inhibits both HIV-1 and HIV-2 proteases. It is a potent inhibitor of HIV replication in vitro and suppresses viral load and increases CD4+ cell counts in HIV patients in vivo. Chemically saquinavir designated as (2S)-N-[(2S,3R)-4- [(3S)-3- (tert-butylcarbamoyl)-decahydroisoquinoline -2-yl] -3-hydroxy -1-phenylbutan -2-yl] -2- (quinolin- 2- ylformamido) butanediamide with a molecular weight and formula of 670.841 and C38H50N6O5(Figure 1).5,6
Literature on saquinavirrevealed that no analytical procedure for the estimation of drug in bulk and formulations was reported on LC-MS/MS. Few methods were reported on spectroscopy,7 liquid chromatography 8-11 and ion pair chromatography.12 The development of specific method like LC-MS/MS is highly essential for the quantification of saquinavir in biological matrices.
MATERIALS AND METHODS
Chemicals and reagents
Saquinavir (99.91% purity) was acquired from Hetero drugs, Hyderabad, India. Methyl indinavir (Internal standard) of 99.89% was received from MSN Labs, Hyderabad, India. Acetonitrile and methanol of HPLC-grade and formic acid of analytical grade were procured from J.T.Baker, Hyderabad, India. The water of Milli-Q was utilized during the processing of the movable phase and washing solvents, acquired from the in-built Milli-Q®RO system.
Liquid chromatographic-MS/MS system
An LC-MS/MS SCIEX API4000 instrument furnished with a positive Electro-spray ionization (ESI) source and Shimadzu prominence HPLC comprising of SIL-HTC autosampler, binary pumps and column oven was utilized in the present work. Analyte quantitation, acquisition of data and it’s integration were processed by utilizing analyst software-1.6.3 version.
Liquid chromatography
Chromatographic elution was attained on a Thermo Hypersil ODS stationary column having the dimensions of 50 × 4.6 mm and a particle size of 2.4 μm. The column oven temperature was monitored at 30.0 °C and 0.5 mL/min flow rate. The injection volume was 5 μL. Isocratic elution was processed with methanol and 0.1%V/V formic acid in the ratio of 90:10 V/V as the mobile phase. Saquinavir and indinavir internal standard were separated in a total runtime of 6 minutes. The autosampler temperature and analytical Column temperatures were kept at 5.0 °C and 30.0 °C respectively.
Mass system conditions
The mass spectrometer was operated under multiple reactions monitoring (MRM) mode in positive ion mode for both saquinavir and internal standard (IS). The adjusted mass system parameters for saquinavir and IS were: both of drying gas and sheath gas temperatures were 350 °C. Nebulizer pressure was monitored at 25.0 psi. The sheath gas and drying gas flow rates were 10.0 L/min and 5.0 L/min, respectively. Capillary voltage was set at 3.0 kV. Collision energy and fragment or voltage were 15 eV and 120 V for saquinavir, 15 eV and 100 V for IS. The transitions (precursor to product) monitored were m/z 671.3 → 654.3 for saquinavir and 614.3 → 596.3 for IS. The dwell time was 200 ms for each transition.
Calibration standards
A stock solution of 1mg/mL solution of saquinavir was prepared freshly by dissolving 10mg of drug in 10mL of 70% methanol. Calibration standards of eight different concentration levels were processed by spiking blank plasma with saquinavir standard solution to get the concentrations of 260.4, 520.8, 1302, 3100, 5100, 6950, 8750 and 10416 ng/mL.
Quality Control Standards
These solutions were prepared at three different levels of lowest quality control(LQC), standards, median quality control(MQC) standards and highest quality control(HQC) standards. These quality control(QC) samples were prepared according to calibration standards to get the concentration of 729.1, 5208 and 7812 ng/mL for LQC, MQC and HQC respectively. Prepared solutions were store at -200c till the time of analysis.
Sample Preparation Method
Sample solution was prepared by transferring 300 µl of plasma and 50µl of indinavir (1µg/mL) into a polypropylene tube and vortexed for 2 min. Saquinavir and IS were extracted with 4mL of ethyl acetate as solvent followed by centrifugation of the solution at 4500 rpm/min for 20min. After centrifugation, organic phase was separated and it was dried using lyophiliser. The resulting product was made solubilize in 250µl of mobile phase and then it transferred into a pre-labeled vials. The vials were kept in an auto-sampler and infused into LC-MS/MS system.
Method validation
The developed method was validated for the parameters: specificity, selectivity, linearity, stability, matrix effect, recovery, precision and accuracy.13, 14
RESULTS
Mass system conditions
For the mass determination, we applied the MRM in positive mode with saquinavir to achieve higher sensitivity and better specificity. ESI was expected to be the best source of ions for LC–MS/MS and the precursor ions and product ions were ascertained by infusing the standard solutions into the mass instrument through a syringe pump. The product ions mass spectra of saquinavir and IS were obtained at m/z 654.3 an 596.3 which were selected as detection ions. Meanwhile, the parameters of mass spectrum environments (capillary voltage, ION-spray voltage and temperature, heater gas, nebulizer gas, collision gas and curtain gas and so on) were optimized to acquire higher mass response.
Internal standard selection
In the present study, indinavir was selected as the IS because it had similar chromatographic behaviors, ionization, extraction efficiency and retention action as the saquinavir and there was no obvious interferences found at the retention times of analytes and IS according to the method validation results.
Method validation
Specificity
Blank plasma samples obtained from six different batches of human plasma was spiked with the saquinavir at LLOQ and IS to assess the specificity. As Figure 2 showed, the retention times of saquinavir and IS were found to be 5.0 and 3.2 min, respectively. There was no noticeable intrusion of matrix substance and internal standard on saquinavir analytical data, and the response of all interfering peaks were less than 20% of LLOQ samples. 14, 15
Sensitivity and linearity
The method linearity was established for saquinavir and excellent linearity was exhibited in the concentration range of 260.4 to 10416 ng/ml (Table 1). The linearity graphs were produced by the peak area ratios (y) of the saquinavir to the IS versus the concentrations (x) with the 1/C2 weighting factor. The developed method linearity equation of the calibration graphs was y = 0.00843x + 0.00249 with a correlation coefficient (r2) value of 0.9991. The LLOQ of saquinavir was 260.4 ng/ml (signal-to-noise ratio > 5.0) assessed by 5 replicate sample solutions which were adequate for reliable quantitation of saquinavir in the plasma samples study.16
Accuracy and precision
The intra-batch and inter-batch precision and accuracy were evaluated by six spiked samples of plasma with saquinavir at LLOQ, LQC, MQC and HQC levels in a batch and in 3 succeeding batches, respectively. The results of precision and accuracy for the quantitation of saquinavir were represented in the Table 2. The percentage relative standard deviation values of inter and intra-batch precision were present in the range of 2.8 to 5.3. The deviation of relative error in the accuracy of inter and intra-batch values were present in between -4.52 to 5.88. 16, 17-21
Extraction recovery
The biological samples were effectively pretreated before the analysis. The extraction recovery was evaluated by calculating the peak area ratio of LQC, MQC and HQC level solutions (n=6) of saquinavir to the extracted spiked samples at respective concentration levels. In same manner, extraction recovery of indinavir was assessed by calculating the peak area ratio of quality control plasma sample solutions (n=6) to spiked human plasma samples at respective concentration levels. The mean extraction recovery of saquinavir was 94.6%, 92.4 and 104.2% at low, medium and high QC levels respectively. The mean extraction recovery of indinavir was 94.8% at a concentration of 100 ng/L. The results were represented in Table 3 and Figures 3 to 5.15, 18-21
Matrix effect
Co-eluting matrix constituents can reduce or increase ionization process in mass system, which may not produce a detectable response in blank matrix due to accuracy of the method. Hence, the different matrix-related component ions dominance was assessed by estimating the IS normalized matrix factor in eight individual sources (consisting of 2-hemolytic and 2-lipemic batches) of human plasma. The average IS normalized matrix factor for all the analytes were present in between 0.95 and 1.06 with a %RSD of ≤4.99 and the results were shown in Table 4. 16
Dilution integrity
The dilution integrity was processed at twice the concentration ULOQ (upper limit of quantification) for saquinavir. After the 1:4 dilution the average back-calculated drug content for dilution quality control samples were within 85–115% of the nominal value with a %RSD of ≤4.5.
Stability
Stability of saquinavir was evaluated in both matrix and aqueous-based samples. Saquinavir and indinavir in-stock solutions were not affected at 1–10°C for 70 days and stock solutions in diluent were not affected at 1–10°C for 48 h. Matrix stability was assessed at −70 and −20°C for 60 days. Matrix stability was analyzed against freshly prepared spiked linearity standards. The findings of stability data were represented in Table 5. Drug was not degraded upto 20 h at the bench-top at a temperature of below 10°C and after six freeze-thaw cycles. The prepared sample solutions were stable at 10°C for 72 h in the auto-sampler. 17
DISCUSSION
Method development was processed with different mobile phases and finally optimized with methanol and 0.1%V/V formic acid in the ratio of 90:10 V/V thru Thermo Hypersil ODS (50 × 4.6 mm, 2.4 μm) column. The optimized method was validated as per the FDA and EMA guidelines for the specificity, selectivity, linearity, stability, matrix effect, recovery, precision and accuracy parameters. The LLOQQC was analyzed with satisfactory accuracy, < 20% of precision and > 5 of signal-to-noise (S/N) ratio. To evaluate the carry-over effect, the blank plasma solution was infused into the chromatographic system after analyzing the highest concentration level of calibration standard in an analytical run and there was no visible carry-over effect was found. The method has linearity in the concentration range of 260.4 to 10416 ng/ml. These findings of precision and accuracy proved that they were reliable for the estimation of saquinavir in human plasma. Method has good extraction recovery with mean recovery of 97.07%. Method was subjected for freeze-thaw, refrigerator, bench-top, long term and in-injector stability and has high degree of stability.
CONCLUSION
In the present research work, a sensitive and accurate LC-MS/MS technique was developed and validated for the successful determination of saquinavir in human plasma. This method exhibited excellent specificity, linearity, accuracy, precision and stability. The linearity equation and correlation coefficient (r2) findings were y = 0.00843x + 0.00249 and 0.9991 respectively. The %RSD of intra and inter-day precision of the developed technique was found between 2.8 to 5.3% for the QC samples (260.4, 729.1, 5208 and 7812 ng/ml). Hence, the validated method can be applied for the pharmacokinetic and toxicokinetic studies in the clinical and forensic analysis of saquinavir in different kinds of biological matrices successfully.
ACKNOWLEDGEMENT
Authors acknowledge immense help received from the scholars whose articles are cited and included in reference to the manuscript. The authors are grateful to all the authors/ editors/publisher of those articles, journals, and books from where the literature of this article has been reviewed and discussed. There is no source of funding for this study. There is no conflict of interest in this study.
Source(s) of funding: No funding is involved.
Conflicts of interest: The authors declare no conflicting interest.
Author’s contribution: All the authors contributed equally.
References:
-
Invirase (saquinavir mesylate) Capsules and Tablets, for Oral Use. Full Prescribing Information (PDF). Genentech, Inc. Archived (PDF) from the original on 24 November 2015. Retrieved 23 November 2015.
-
Saquinavir. The American Society of Health-System Pharmacists. Archived from the original on 2015-09-08. Retrieved Sep 5, 2015.
-
Minor, Lisa K. Handbook of Assay Development in Drug Discovery. Hoboken: CRC Press; 2006. p. 117.
-
Hamilton, Richart. Tarascon Pocket Pharmacopoeia 2015 Deluxe Lab-Coat Edition. Jones & Bartlett Learning; 2015. p. 72.
-
Winston A, Back D, Fletcher C. Effect of omeprazole on the pharmacokinetics of saquinavir-500 mg formulation with ritonavir in healthy male and female volunteers. AIDS. 2006;20(10):1401–6.
-
HIV Surveillance-United States, 1981-2008. Archived from the original on 9 November 2013. Retrieved 8 November 2013.
-
Sankar DG, Kumar JMR, Reddy MVVN. UV spectrophotometric methods for the determination of Saquinavir mesylate and efavirenz. A J of Chem. 2003; 15(3):1856-1858.
-
Riem Ha H, Follath F, Bloemhard Y, Krahenbuhl S. Determination of saquinavir in human plasma by high-performance liquid chromatography. J of Chrom.B. 1997;694:427–433.
-
Ucpinar SD, Stavchansky S. Quantitative determination of saquinavir from Caco-2 cell monolayers by HPLC-UV. High-performance liquid chromatography. Biomed Chromatogr. 2003;17(1):21-25.
-
Takahashi M, Yoshida M, Oki T, Okumura N, Suzuki T, Kaneda T. Conventional HPLC Method Used for Simultaneous Determination of the Seven HIV Protease Inhibitors and Non-nucleoside Reverse Transcription Inhibitor Efavirenz in Human Plasma. Biol. Pharm. Bull. 2005;28(7):1286—1290.
-
Prakash VK, Venkateswara Rao J, Appala Raju N. Estimation of Saquinavir Mesylate in Tablet Dosage Form using RP-HPLC. A J of Chem. 2009;21:612-616.
-
Richard Hoetelmans MW, Marjolijn Van E, Pieter Meenhorst L, Jan Mulder W, Jos Beijnen H. Determination of saquinavir in human plasma, saliva, and cerebrospinal fluid by ion-pair high-performance liquid chromatography with ultraviolet detection. J Chrom. B. 1997; 698:235-241.
-
FDA Guidance for Industry, Bioanalytical Method Validation, US Department of Health and Human Services, Food and Drug Administration, Centre for Drug Evaluation and Research (CDER), Center for Veterinary Medicine (CVM) May 2001.
-
European Medicines Agency, Guideline on bioanalytical method validation 2011.
-
Henion J, Brewer E, Rule G. Sample Preparation for LC/MS/MS: Knowing the Basic Requirements and the Big Picture of an LC/MS System can Ensure Success in Most Instances. Anal Chem. 1998;70:650A-656A.
-
Chambers EE, Woodcock MJ, Wheaton JP. Systematic development of an UPLC–MS/MS method for the determination of tricyclic antidepressants in human urine. J. Pharm. Biomed. Anal. 2014;88:660–665.
-
Patel DS, Sharma N, Patel MC. Development and validation of a selective and sensitive LC–MS/MS method for determination of cycloserine in human plasma: application to bioequivalence study, J Chromatogr B. 2011; 879:2265–2273.
-
Hindu K, Vinodhini C, Srinivas SK, Rajan SM, Chitra K, Mangathayaru K. Validated RP-HPLC Method for Quantification of Paclitaxel in Human Plasma – Eliminates Negative Influence of Cremophor EI. IJCRR. 2018;10(13):5-10.
-
Murphy AT, Kasper SC, Gillespie TA, DeLong AF. Determination of Xanomeline and Active Metabolite, N-Desmethylxanomeline, in Human Plasma by Liquid Chromatography-Atmospheric Pressure Chemical Ionization Mass Spectrometry. J. Chromatogr B. Biomed Appl. 1995;668:273-280.
-
Nagalakshmi V, Srinivas Rao G, Gayathri Devi N, Mohan S. RP-HPLC Method for Simultaneous Estimation of Vildagliptin and Metformin in Bulk and Pharmaceutical Formulations. IJCRR. 2021. 13(7)112-117.
-
Cheruku S, Bhikshapathi DVRN. Validation of a specific LC-MS/MS method for the quantification of anti-cancer alectinib in biological matrices. IJPR. 2021;13(1):6513-21.
|