An In Vitro Assessment of Retention Force for Implant-Retained Overdentures Using All-Polyetheretherketone, Zirconia, and Titanium Attachments
Author: Mahalakshmi Gujjalapudi, Amit Kumar Verma, Swaroopkumar M Magar, Doddy Lokanathan Balaji, Gagandeep Singh Dang, Atul Bhardwaj
Category: Healthcare
Abstract:Introduction: Numerous problems were described by complete denture wearers for example lack retention and stability of these dentures, thus decreasing chewing efficacy, and pain during mastication. These problems can be effectively overwhelmed with implant-assisted overdentures. Objectives: This study was done to assess the retention forces of all-zirconia (ZrO2 ), all-polyetheretherketone (PEEK), and titanium for two implant-retained mandibular overdentures. Materials and Methods: In each model, twenty-four similar acrylic resin mode positioned at the canine area. Models were categorized into 4 groups of 7 samples in each group; all all-zirconia (ZrO2 ), all-polyetheretherketone (PEEK), titanium and a combination of PEEK with zirconia. Alike experimental overdentures were prepared for all models. A universal testing machine was used for assessing the retention force, using the pull-off test in the existence of artificial saliva in-between the crowns. A higher score of retention force was noted at the beginning and after 540 cycles of insertion and elimination which mimicking 6 months of the clinical outcome of final retention. Obtained results were statistically evaluated. Results: Zirconia group conveyed the greatest initial force values. A combination of zirconia and PEEK was found to be effective. The final retention values of the PEEK and Zirconia groups were significantly decreased than initial values, while the insignificant loss of retention was observed with the Titanium group (P = 0.05). Conclusion: Zirconia (ZrO2 ) with PEEK combination is more preferred as a secondary telescopic. Despite the retention loss with time, even PEEK can be recommended as a secondary telescopic crown beside PEEK or ZrO2 primary crowns concerning the adequate early and final retention scores.
Keywords: PEEK, Retention force, Telescopic attachment, Zirconia
Full Text:
INTRODUCTION
Numerous problems were described by complete denture wearers for example lack retention and stability of these dentures, thus decreasing chewing efficacy, and pain during mastication. These problems can be effectively overwhelmed with implant-assisted overdentures.1
Resilient telescopic supplements have been used since 1989, to retain overdentures for rehabilitation of the edentulous mandible. In cases of overdenture for long term success, two interforaminal implants with resilient telescopic attachments appeared to be an effective and efficient treatment option in cases of the severely atrophied edentulous mandible.2
In 2000 using ceramic materials for the fabrication of telescopic attachments was first defined. Zirconia (ZrO2) is a ceramic material that is largely used in current years for medical devices, displaying excellent mechanical strength, high biocompatibility, and wear resistance. The use of tooth-coloured ceramic materials display a positive psychological outcome on patients and upholds improvements in oral hygiene.3
Polyetheretherketone (PEEK) is a thermoplastic polymer, comprise of a molecular chain from an aromatic backbone that is connected by ketone and ether function group.4 PEEK exhibits high biocompatibility and can be used for numerous dental conditions, such as dental implants, and provisional abutments.5,6
The present study was done to evaluate the retention force of different CAD/CAM telescopic attachments (zirconia, PEEK, and titanium telescopic attachments) for retention of mandibular implant overdenture.
MATERIALS AND METHOD
In the present study twenty-one, acrylic resin models of edentulous mandible without alveolar undercuts were used. Every model was surrounded by a 2 mm layer of silicon soft liner material to mimic alveolar mucosa in accordance to El-Charkawi et al.7 Two implants with 10 mm length and 4.5 mm diameter (dentium implants) were mounted in the canine area of the respective model with the help of a template guide,8 later two dual abutments with 4.5 diameters, 1.5 gingival height and 4.5 lengths, were positioned into the implants. For overdenture construction, the resin models were duplicated. Models were divided into 4 groups; group zirconia, PEEK, titanium and PEEK+ zirconia materials.
Fabrication and designing and of telescopic attachments
Primary resilient telescopic crown construction was done with the succeeding parameters for all groups; 5 mm height (2 mm gingival height was paralleled and the occlusal 3 mm was tapered 4°). The primary crowns were intended to ensure a shared path of insertion. The computer numeric control data were preserved as STL files. From semi-sintered ZrO2 blanks (ZrO2, Kataya) the primary ZrO2 crowns were milled and from BioHPP blanks (Bredent, UK) the primary PEEK crowns were milled. Using zinc phosphate cement on an abutment the primary crowns were cemented.
Secondary resilient telescopic crown construction was done by scanning each model, followed by separate scans of each primary crown to improve the quality of data. The subsequent parameters were employed for designing secondary crowns; parallel walls with a minimal wall thickness of 0.5 mm and occlusal space (0.3 mm) built-in between the primary and secondary crowns.2 Mechanical projections were added to the design of each secondary crown, and later secondary ZrO2 and PEEK crowns were adopted.
Secondary crown procedures
The fitting area was altered by creating vent holes through the lingual flanges of each overdenture. Secondary crowns were built-in over the primary ones in the correct path of insertion and then picked up to the fitting surface of the overdenture using an auto polymerized acrylic resin.
Measurement of retention force
With auto polymerized acrylic resin a metallic cobalt chrome bar (2 mm, 15 mm, and 120 mm), with a grasping hook in the middle, was covered. 9 For measuring the retention force, models were placed and fixed in a universal testing machine (LLOYD instruments) then a pull-off test was performed with a load cell of 3.5 KN and a crosshead speed of 50 mm/min. Overdenture with each combination of secondary and primary telescopic crowns was exposed to 540 cycles of enclosure and exclusion representing 6 months of clinical service under the same circumstances. Then, the pull-off test was achieved again to assess the final retention.
The obtained data were tabulated and statistically evaluated using the SPSS version 21(SPSS Inc., Chicago, Illinois, USA) using Shapiro–Wilk test, Independent samples test, ANOVA was used followed by the post hoc Tukey test. P-value was lesser than 0.05 measured as statistically significant.
RESULTS
Table 1 indicated, the average values of initial and final retention values (telescopic attachment) in all groups. The highest mean initial retention value for all groups was observed in Group-I where the mean initial retentive force was 21.2550 N and the final one was 16.1140 N. However, mean initial retentive forces of group 2 (Titanium) and 3 (PEEK) 14.6310 N and 15.3530 N respectively and their final retentive forces were 14.4160N and 14.0470 N respectively. Group IV for the combination of Zirconia+ PEEK showed initial forces as 17.3430 N and 16.0340 N final forces.
Comparison of mean values between initial and final retention values within each group using paired t-test showed that the final retention value for the PEEK and Zirconia and Zr+PEEK was significantly decreased by time (P = 0.0001, 0.0000 and 0.0001) respectively), but no significant loss in retention force value was observed for the titanium group where P = 0.057.
Table 2 indicated the Initial and final retention values comparison among all groups. There was a significant variance in both initial and final retentions between every two groups using the Independent samples test.
DISCUSSION
For achieving clinical performance and patient satisfaction of the prosthesis, it is necessary to maintain retention of the attachment system.3
In the present study, the highest mean initial retention values were found in the Zirconia group followed by PEEK. Our results are in agreement with the in vitro study of Schubert et al.;10 they observed that CAD-CAM fabricated PEEK secondary crowns can deliver sufficient retention force values.
Final retention values of PEEK and Titanium groups were significantly decreased than initial values, while insignificant loss of retention was observed with the zirconia group; this may be owing to wear and alteration in surface topography of the secondary and primary crowns opposing areas. It was described that, during the function of telescopic attachment, frictional wear can occur.11,12
Emera et al., assessed the changes in surface topography of all ZrO2, ZrO2-PEEK and all PEEK, telescopic attachments after simulated 6 months of using the overdenture.13 They determined that the PEEK and ZrO2 combination for the fabrication of telescopic attachment was attended with superior variations in surface topography (mostly in secondary crowns) in association with all PEEK and all ZrO2 telescopic connections.14, 15
A significant change among the initial and final retention between all groups was observed due to significant wear that occurred in all groups after simulating 6 months of overdenture.16
CONCLUSION
It can be concluded that ZrO2 is further desired as a secondary telescopic crown. Although the loss of retention over time, PEEK can still be used as a secondary telescopic crown against titanium or PPEK primary crowns regarding the acceptable initial and final retention values.
Conflict of interest: nil
Source of funding: self
Authors contribution
-
Dr. Gujjalapudi M- Editing
-
Dr. Amit Verma- Manuscript writing
-
Dr. Swaroopkumar M Magar- Data collection, investigation
-
Dr. Balaji- Review
-
Dr. Gagandeep Singh Dang- Review
-
Dr. Atul Bhardwaj- Editing
![](https://ijcrr.com/admin/public/uploads/1/myfiles/1_1640429968341.png)
References:
-
Chen YF, Yang YH, Lee JH, Chen JH, Lee HE, Chou TM. Tongue support of complete dentures in the elderly. Kaohsiung J Med Sci. 2012;28:273-278.
-
Heckmann SM, Schrott A, Graef F, Wichmann MG, Weber HP. Mandibular two-implant telescopic overdentures. Clin Oral Implants Res. 2004;15:560-569.
-
Weigl P, Hahn L, Lauer HC. Advanced biomaterials used for a new telescopic retainer for removable dentures. J Biomed Mater Res. 2000;53:320-336.
-
Kurtz SM, Devine JN. PEEK biomaterials in trauma, orthopaedic, and spinal implants. Biomater. 2007;28:4845-4869.
-
Merk S, Wagner C, Stock V, Eichberger M, Schmidlin PR, et al. Suitability of secondary PEEK telescopic crowns on zirconia primary crowns: The influence of fabrication method and taper. Materials. 2016;9:908-917.
-
Stock V, Wagner C, Merk S, Roos M, Schmidlin PR, et al. Retention force of differently fabricated telescopic PEEK crowns with different tapers. Dent Mater J. 2016;35:594-600.
-
El-Charkawi H, Zekry K, Elwaked M. Stress analysis of different osseointegrated implants supporting distal extension prosthesis. J Prosthet Dent. 1994;72: 1-5.
-
Ebadian B, Talebi S, Khodaeian N, Farzin M. Stress analysis of mandibular implant-retained overdenture with independent attachment system: Effect of restoration space and attachment height. Gen Dent. 2015;63:61-67.
-
El Mekawy N, Khalifa A, Abdualgabbar E. The influence of palatal coverage on the retention force and fatigue resistance of mini dental implant maxillary overdenture. J Oral Hyg Health. 2016;4:200-208.
-
Schubert O, Reitmaier J, Schweiger J, Erdelt K, Güth JF. The retentive force of PEEK secondary crowns on zirconia primary crowns over time. Clin Oral Investig. 2019;23:2331-2338.
-
Bayer S, Kraus D, Keilig L, Gölz L, Stark H, Enkling N. Wear of double crown systems: Electroplated vs. casted female part. J Appl Oral Sci. 2012;20:384-391.
-
Majcher A, Le?niewska-Kochanek A, Mierzwi?ska-Nastalska E. A method and a device for the evaluation of the retention of telescopic dental crowns. J Mech Behav Biomed Mater. 2017;69:362-367.
-
Emera R, Elgamal M, Albadwei M. Surface wear of all zicronia, all PEEK and zirconia-peek telescopic attachments for two implants retained mandibular complete overdentures. In -vitro study using the scanning electron microscope. IOSR J Dent Med Sci. 2019;18:59-68.
-
Turp I, Bozda? E, Sünbülo?lu E, Kahruman C, Yusufo?lu I, Bayraktar G. Retention and surface changes of zirconia primary crowns with secondary crowns of different materials. Clin Oral Investig. 2014;18:2023-2035.
-
Rutkunas V, Mizutani H, Takahashi H. Influence of attachment wear on retention of mandibular overdenture. J Oral Rehabil. 2007;34:41-51.
-
Lughi V, Sergo V. Low-temperature degradation – ageing – of zirconia: A critical review of the relevant aspects in dentistry. Dent Mater. 2010;26:807-820.
|